Zahn Cup Sizes Test
The Zahn Viscosity Cup is used to measure and adjust the viscosity of ink. The viscosity will be measured by the time the ink took to flow down from the orifice.
Zahn Cup Sizes Test
Cup No | Viscosity※ | Time |
1 | 20 to 85 cP | 40 to 85 s |
2 | 30 to 170 cP | 20 to 70 s |
3 | 170 to 550 cP | 15 to 60 s |
4 | 200 to 900 cP | 20 to 65 s |
5 | 250 to 1,200 cP | 15 to 60 s |
BGD 126 Zahn Cups can be used to quickly measure the viscosity of liquids such as Paint, Ink, Varnishes, Syrup and Oil. They are produced according with ASTM D 816, ASTM D 1084,and ASM D 4212. The Stainless Steel Cups are molded precisely and the orifices are precision drilled using high-speed CNC drilling.
Each cup has a 12-inch loop handle to allow the cup to be dipped by hand into a liquid container. At the center of this handle is a finger-ring for holding the cup in a vertical position during use. Their orifice diameters are set at the factory for appropriate results with applicable NIST traceable Newtonian oils.
It is done by filling up the volume of Zahn Cup with the required liquid and then placing an index finger over the orifice to stop any flow out of the liquid. A ring at the Zahn Cup handle allows the cup to be hanged perpendicularly(still with the finger stopping any liquid flow out). Prepare a Digital stopwatch and set zero. Immediately start the stopwatch when the finger is released. As soon as there is a first “break” of the flow, stop the stopwatch. The second shown is the time of the liquid flow, which is also the flow viscosity of that liquid.
Every Biuged’s Zahn cup has its own coefficient “K”, which is marked on the cup body.
Fact Viscosity = Testing Viscosity × K
Main Technical Parameters
- Volume of Cup :43ml - 49ml
- Length of Handle:40mm ± 0.1mm
- Height of Cup:58mm ± 0.1mm
- Measurement Temperature:25℃±1℃
- Come with calibration certificate
- Details of each Zahn Cups
- Click Here : Zahn Cup / Zahn Viscosity Cup Brochure
Zahn Cup Sizes Test
Details / Zahn Cup |
Zahn Cup No.1 | Zahn Cup No.2 | Zahn Cup No.3 |
Orifice (mm/inches) |
1.98 / 0.08 | 2.74 / 0.11 | 3.76 / 0.15 |
Flow Time Range (sec) |
33.5 - 80 | 20 - 80 | 20 - 75 |
Measurement Range (CST) |
5 - 56 | 21 - 231 | 146 - 848 |
Applications | Thin-bodied materials | Clears, lacquers, enamels, press-side adjustment of flexographic inks |
Manufacturing of flexographic inks |
Ordering Information |
BGD 126/1 | BGD 126/2 | BGD 126/3 |
Zahn Cup Sizes Test | |||
Details / Zahn Cup |
Zahn Cup No.4 | Zahn Cup No.5 | |
Orifice (mm/inches) |
4.27 / 0.17 | 5.28 / 0.21 | |
Flow Time Range (sec) |
20 - 80 | 20 - 80 | |
Measurement Range (CST) |
222 - 1110 | 460 - 1840 | |
Applications | More viscous paints and inks |
Silk screen inks | |
Ordering Information |
BGD 126/4 | BGD 126/5 |
Zahn Cup Sizes Test
Untuk informasi,pertanyaan dan pemesanan silahkan hubungi kami di :
- PT. Mora Anugerah Berkat [M.A.B]
- Click Here : Zahn Cup / Zahn Viscosity Cup Brochure
- Alamat : Jl.Taman Asri Blok M No.36 Kel.Cipadu Kec.Larangan, Kota Tangerang - Provinsi Banten
- Phone : 021 7345.7064
- Mobile : 0812.8140.0749 / 0812.8770.5470
- Chat : Whats App, Line, WeChat, Telegram
- Website : www.moraanugerah.id - www.moraanugerah.com
- Email : info@moraanugerah.com - marketing@moraanugerah.com -
- marketing_div@moraanugerah.com