Blog


Nippon - Mercury Analyzer RA-5A

5 November 2024 ENVIROMENTAL TESTING BLOG

PT.Mora Anugerah Berkat (M.A.B) merupakan distributor/supplier dari alat-alat laboratorium khususnya dibidang laboratorium lingkungan, kami menyediakan alat pengujian lingkungan diantaranya adalah NIC - Mercury Analyzer RA-5A

The RA-5A delivers ultra-high sensitivity and excellent precision while only requiring 5ml os sample. The RA-5A is also environmentally-friendly thorughits direct purge design that reduces reagent consumption and waste production by over 80% versus comparable model without any need for compressed gases RA-5A has achieved the highest level sensivity and accuracy with new innovatory-crafted detector optics.

NIC - Mercury Analyzer RA-5A

Sample N Average C.V
0.05μg/L 5 0.052μg/L 2.2 %
0.10µg/L 5 0.101μg/L 1.70%

Click Here : NIC Mercury Analyzer RA-5A Brochure

Nippon Instruments Corporation
Specifications Mercury Analyzer RA – 5A

Data

Processing

Software

Measurement Principle Non-dispersive double beam cold vaporatomatic absorption spectrometry [Patent Pending]
Measurement Method Reducing Vaporrization with open air-supply (Peak height measuring and integration measuring salectable)
Light Source Low pressure mercury discharge lamp
Wave Length 253.7 nm
Detector semicondutor sensor (w/254nm bandpass filter)
Detection Limit 0,0025ng/5ml (0.5ppt)
Maximum Range 2.000ng/5ml (400 ppb)
Carrier Gas Required No0 Compressed purified gas Required
Measure Flow Rate Adjustable (0.1 - 0.9 L/min) (just operet with ambient air)
Measuring Time 30sec - 999sec
Dehumidifier Peltier device
Mercury Removal Removal by activated carbon

Mercury

Detection

Section

OS Windows 10 Pro
Comunication USB
Display Peak Wave shape,calibration curve,measuring time and measuring result
Data Selectable from 6 formula (3 liniear and 3rd order fits), Correlation
Processing coefficient is automatically calculated Static calculation : Mean Value,standard deviation Coefficient of variation
Software (All value are automatically calculated), self check function
Print Measuring conditions,Calibration Curve,Measured values,Statical Calculations Peak wave shape and Measuring time
Quality Control Check Standard, Avoid abnormally high mercury
Instalation Dimension 248W x 253D x 180H (mm)
Weight 4,7 (Kg)
Instalation Enviroment 10°C - 35°C, < 10 - 80% RH (Non condensing enviroment is allowed)
Power Supply AC, single phase, 100 - 240V 50/60 Hz, 150VA (AC Adaptor Included)

NIC - Mercury Analyzer RA-5A

Kami juga menyediakan Jasa Perbaikan (SERVICE), Jasa Perawatan (Maintenace) dan support untuk KALIBRASI / AKSESSORIES SPAREPART jangan ragu untuk menghubungi kami marketing dan sales kami siap membantu menjawab segala kebutuhan anda.

  • PT. Mora Anugerah Berkat (M.A.B)
  • Click Here : NIC Mercury Analyzer RA-5A Brochure
  • Alamat : Jl.Taman Asri Blok M No.36 Kel.Cipadu Kec.Larangan,Kota Tangerang - Provinsi Banten
  • Phone : 021 7345.7064
  • Mobile : 0812.8140.0749 / 0812.8770.5470
  • Chat: Whats App,Line,WeChat,Telegram
  • Website : www.moraanugerah.id - www.moraanugerah.com
  • Email : info@moraanugerah.com - marketing@moraanugerah.com - marketing_div@moraanugerah.com