Ventilator
PARAMETER | SPECIFICATION | |
Display Mode | High-definition 5.7” LCD screen display | |
Ventilation Mode | IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH | |
Ventilation |
Tidal Volume | 50~1500ml |
Rate | 2~99bpm | |
SIMV Rate | 2~20bpm | |
I:E | 2:1~1:8 | |
Inspiratory Trigger Pressure | -1.0~2.0kPa | |
PEEP | 0~2.0kPa | |
Pressure Range | 1.0~6.0kPa | |
SIGH | 1.5 times the inspiratory time | |
Parameters for Ventilation Monitoring |
Tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate,I/E, peak pressure of airway, pressure - time waveform, flow rate – time waveform,PEEP, inspiratory trigger pressure. | |
Power | AC 220V±10% 50Hz±1Hz; UPS, Storage battery | |
Mechanical Arms | ||
Power Alarm | SECURITY ALARM SYSTEM | |
Airway Pressure |
Upper Limit Setting Range | 1.0~6.0kPa |
Low Limit Setting Range | 0.4~2.0kPa | |
Per-minute Ventilation |
Upper Limit Setting Range | 3.0~30L/min |
Low Limit Setting Range | 1.0~10L/min | |
Sustained High-pressure Alarm |
It will give alarm when stress have consistently been higher than 2.5kPa. |
|
Suffocation Alarm | It will give sound and light alarm if there is no tidal volume input for 15 seconds. |